OSCE Practice – Autoclave & Washer-Disinfector Testing
Practise recognising and performing equipment testing and validation procedures in a dental setting, following NEBDN OSCE expectations.
Scenario
You are asked to explain the daily, weekly, and annual checks required for autoclaves and washer-disinfectors.
Your Task
- Identify required testing procedures for equipment
- Perform or recognise daily and weekly tests
- Ensure correct use of autoclaves
- Record cycle and test results accurately
- Maintain compliance with HTM 01-05 and CQC standards
Please log in to access this Quiz.
Key Learning Points
- Daily autoclave testing includes vacuum test and automatic control test (ACT)
- Weekly washer-disinfector testing includes protein residue test
- Type B autoclaves are used for wrapped, hollow, and complex instruments
- Cycle parameters (time, temperature, pressure) must be recorded after each cycle
- Annual validation by a competent engineer is required for compliance